|
|
|
|
Non
Contact AFM mode
|
|
|
|
Figure
1 compares the movement of the probe tip relative to the sample surface for
images being acquired between in contact AFM and in non-contact AFM. Contact AFM
uses the physical contact between the probe tip and the sample surface, whereas
non-contact AFM does not require this contact with the sample. In Non-Contact
mode, the force between the tip and the sample is very weak so that there is no
unexpected change in the sample during the measurement. Therefore, Non-Contact
AFM is very useful when a biological sample or other very soft sample is being
measured; the tip will also have an extended lifetime because it is not abraded
during the scanning process. On the other hand, the force between the tip and
the sample in the non-contact regime is very low, and it is not possible to
measure the deflection of the cantilever directly. So, Non-Contact AFM detects
the changes in the phase or the vibration amplitude of the cantilever that are
induced by the attractive force between the probe tip and the sample while the
cantilever is mechanically oscillated near its resonant frequency.
|
|
|
 |
| Figure
1. Concept diagram of Contact mode and Non - Contact mode
|
|
|
Keff=
k0 - F¡
(1)
|
|
|
|
When the attractive force is applied, keff becomes smaller than ko
since the force gradient is positive. Accordingly, the stronger the
interaction between the surface and the tip (in other words, the closer the
tip is brought to the surface), the smaller the effective spring constant
becomes. This alternating current method (AC detection) makes more sensitive
responds to the force gradient as opposed to the force itself. Thus, it is
also applied in such techniques as MFM (Magnetic Force Microscopy) and Dynamic
Force Mode AFM. |
|
|
|
A bimorph is
used to mechanically vibrate the cantilever. When the bimorphs drive frequency
reaches the vicinity of the cantilevers natural/intrinsic vibration frequency
(f0), resonance will take place, and the vibration that is
transferred to the cantilever becomes very large. This intrinsic frequency can
be detected by measuring and recording the amplitude of the cantilever
vibration while scanning the drive frequency of the voltage being applied to
the bimorph. Figure 2 displays the relationship between the cantilevers
amplitude and the vibration frequency. From this output, we can determine the
cantilevers intrinsic frequency. |
|
 |
| Figure
2. Resonant Frequency
|
|
On
the other hand, the spring constant affects the resonant frequency (f0)
of the cantilever, and the relation between the spring constant (k0)
in free space and the resonant frequency (f0) is as in Equation
(2).
f0
= (k0 / m)1/2
(2)
As
in Equation (1), since keff becomes smaller than k0 due to the
attractive force, feff too becomes smaller than fo as shown in Figure 3 (a).
If you vibrate the cantilever at the frequency f1(a little larger
than fo) where a steep slope is observed in the graph representing free space
frequency vs. amplitude, the amplitude change at f1 becomes very
large even with a small change of intrinsic frequency caused by atomic
attractions. Therefore, the amplitude change measured in f1
reflects the distance change between the probe tip and the surface atoms.
|
 |
| Figure
3. (a) Resonant frequency shift (b) Amplitude vs Z-feedback |
|
| If
the change in the intrinsic frequency resulting from the interaction
between the surface atoms and the probe or the amplitude change at a given
frequency (f1) can be measured, the non-contact mode feedback
loop will then compensate for the distance change between the tip and the
sample surface as shown in Figure 3 (b). By maintaining constant
cantilevers amplitude (A0) and distance (d0),
non-contact mode can measure the topography of the sample surface by using
the feedback mechanism to control the Z scanner movement following the
measurement of the force gradient represented in Equation (1).
|
|
|
| True
NC-AFM
NC-AFM on biological samples
NC-AFM vs. Tapping mode |


|
|